ANALISIS PENGULANGAN (REPEAT) RADIOGRAFI PADA MODALITAS DIGITAL RADIOGRAFI (DR) DI INSTALASI RADIOLOGI RSU PKU MUHAMMADIYAH BANTUL
Keywords:
Repeat Radiography, Digital Radiography, Contributing Factors, Quality AssuranceAbstract
Background: Digital Radiography (DR) is a digital imaging system that replaces conventional film with digital detectors to produce faster and more efficient diagnostic images. Although this technology has the ability to correct contrast and density errors, repeat examinations still occur due to patient positioning errors, artifacts, and other technical factors. Based on preliminary data from the Radiology Installation of PKU Muhammadiyah Bantul Hospital, image repeats have never been thoroughly analyzed, even though repeats affect radiation dose increases and radiographers' work efficiency.
Objective: This study aims to determine the percentage level of repeat radiography and identify the factors causing repeat examinations using the Digital Radiography (DR) modality at the Radiology Installation of PKU Muhammadiyah Bantul Hospital.
Method: The study used a qualitative approach with a case study design. Data were collected through direct observation, interviews with three radiographers, and documentation of examination data during April–June 2025. Interview data analysis included interview transcripts, data reduction, coding, categories, and themes, followed by conclusions. The repeat results were analyzed descriptively by calculating the percentage of repeats and grouping the causes of repeats based on technical and human error categories. The results were used to assess the implementation of Quality Assurance (QA) and Quality Control (QC) programs in the radiology installation.
Results: Of the total 4,835 radiographic examinations, 35 cases of repeat cases (0.73%) were found, which is below the maximum limit of ≤2% as stipulated in KMK No. 129/Menkes/SK/II/2008. The main factors causing repeat cases were cut objects (0.44%) and artifacts (0.38%), followed by minor cases due to cassettes not being ready (0.12%). Although most radiographers understand the theory and working principles of DR, there is still a mismatch between theory and practice in patient positioning and pre-exposure examination.
Conclusion: The results of the study showed that the repeat rate of radiography at PKU Muhammadiyah Bantul General Hospital was classified as very good (0.73%), indicating the implementation of effective and efficient radiology quality management. These results confirm that radiographers have implemented the principles of QA, QC, and ALARA (As Low As Reasonably Achievable) well, thus reducing repeats below national standards. The main error factors come from human error aspects such as patient positioning and metal artifacts, not equipment damage. Regular evaluation, technical training, and increased accuracy in patient positioning are needed to maintain image quality and patient safety continuously.
References
American Society of Radiologic Technologists. (2024). Best Practices in Digital Radiography. ASRT White Paper. Retrieved from https://www.asrt.org/docs/default-source/research/whitepapers/best-practices-in-digital-radiography.pdf
Anggraeni, D. A., Astari, F. M., & Wiyono, A. B. (2021). Analisa faktor kegagalan foto panoramik dengan studi literatur. Universitas ‘Aisyiyah Yogyakarta. https://digilib.unisayogya.ac.id/6004
British Journal of Radiology, 72(SEP), 856- 863.KEMENKES. 20
Bushong, Steward C. 2017. Radiologic Science for Technologist. Eleventh Edition. Elsivier, inc. Amerika.
Bwanga, O. (2021). Causes of reject and repeat of digital radiographic images: A literature review to guide the practice of radiography in Zambia. Medical Journal of Zambia, 48(1). Retrieved from https://www.researchgate.net/publication/350687601
Chafidh, (2018). Analisis Pengulangan Citra Digital Dengan Menggunakan Digital Radiography Di Instalasi Radiologi Rumah Sakit Panti Waluyo Surakarta.
Delis. 2017. Comparison of film, hard copy digital radiography (DR) and soft copy picture narchiving and communication (PACS) systems using a contrast detail test object.
Fajarrissetyo, I. J., Nurcahyo, P. W., & Aryani, A. I. (2015). Analisis Penolakan dan Pengulangan Citra Radiografi pada Modalitas Computed Radiography AGFA CR 35-X di Instalasi Radiologi RSUD DR. R. Goeteng Taroenadibrata Purbalingga. Jurnal Imejing Diagnostik (JImeD), 1(2), 78–81. https://doi.org/10.31983/jimed.v1i2.3159
Fajarrissetyo, I. (2015). Analisis Penolakan dan Pengulangan Citra Radiografi pada Modalitas Computed Radiography AGFA CR 35-X di Instalasi Radiologi RSUD DR. R. Goeteng Taroenadibrata Purbalingga. J urnal Imejing Diagnostik (JImeD), 1(2), 78-81.
Falentin, F., Astari, F. M., & Za’im, M. (2024). Identifikasi faktor penyebab pengulangan citra radiografi sebagai upaya menjaga keselamatan pasien di Instalasi Radiologi Rumah Sakit Umum ‘Aisyiyah Ponorogo. Journal of Innovation Research and Knowledge, 5(4). https://bajangjournal.com/index.php/JIRK/article/view/11228
Hasaneen, M., et al. (2023). Assessment of image rejection in digital radiography. Radiography, 29(4), 781–789. Retrieved from https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10375339/
Hidayat, R. (2019). Pengaruh pelaksanaan SOP perawat pelaksana terhadap tingkat kecemasan pasien di rawat inap RSUD Bangkinang. Jurnal Ners. https://doi.org/10.3329/bjdre.v4i2.20255
Isnaini, M., Afgani, M. W., Haqqi, A., & Azhari, I. (2025). Teknik Analisis Data Uji Normalitas. J-CEKI: Jurnal Cendekia Ilmiah, 4(2), 1377–1384. Iswanto, H., Titien S, I., & Rahardjo, R. (2015). Penatalaksanaan Impaksi Kaninus Kiri Atas dengan Posisi Horisontal pada Anak. Majalah Kedokteran Gigi Klinik, 1(2), 92. https://doi.org/10.22146/mkgk.11961
Iqbal, M. I. (2021). Analisis faktor penyebab pengulangan foto pada pemeriksaan gigi intra oral. Fakultas Kesehatan dan Keteknisian Medis, Universitas Widyagama Husada Semarang. https://eprints.uwhs.ac.id/91
Journal Center. (2024). Analisis Faktor Penyebab Pengulangan Foto pada Modalitas Digital Radiography di RSUD Dr. Soehadi Prijonegoro Sragen. Jurnal Kesehatan Klinik, 10(2). Retrieved from https://journalcenter.org/index.php/klinik/article/view/3929
Kementrian Kesehatan Republik Indonesia. (2009). Keputusan Menteri Kesehatan Republik Indonesia Nomor 1250/Menkes/SK/XII/2009 tentang Pedoman Kendali Mutu (Quality Control) Peralatan Radiodiagnostik. Jakarta: Kementerian Kesehatan Republik Indonesia.
Kementrian Kesehatan Republik Indonesia. (2008). Keputusan Menteri Kesehatan Republik Indonesia NomoR129/Menkes/SK/II/2008 tentang Standar Pelayanan Minimal Rumah Sakit. Jakarta: Kementerian Kesehatan Republik Indonesia.
Manurung, D., Utama, H., & Rosidah, S. (2014). Analisa Penolakan Foto Gigi Intra Oral di Instalasi Radiologi Rumah Sakit Umum Daerah Muntilan Kabupaten. Magelang. Jurnal Ilmu Dan Teknologi Kesehatan, 5, No 1(2086-8510 (Print), 2655-5875 (Online)), 5–8. https://doi.org/https://doi.org/10.33666/jitk. v5i1
Papp, Jeffrey, PhD, R. (2019). Quality Menagement In The Imaging Science, Sixth Edition. 384.
Presotto, N. D., Teixeira, F. C., Morosolli, A. R. C., & Rockenbach, M. I. B. (2020). Errors in the periapical radiographic technique. International Journal of Applied Dental Sciences. Maxillofacial Journal, 4(02), 104–112.
Riolina, A., & Yulia Indarti, G. (2021).Gambaran Pengetahuan Mahasiswa Non Fakultas Kesehatan Description of Non Health Faculties Student Knowledge About Gingivitis. Biomedika,13(1), 19– 26.
Sari, A. N. V., Yusnida, A. M., & Fakhrurreza, M. (2025). Analisis pengulangan foto rontgen pada modalitas digital radiography di Instalasi Radiologi RSUD Muntilan. Prepotif: Jurnal Kesehatan Masyarakat, 9(2). https://journal.universitaspahlawan.ac.id/index.php/prepotif/article/view/49081
Shrivastava, A., et al. (2018). Reject Analysis in Digital Radiography: A Prospective Study. Indian Journal of Applied Radiology Studies (IJARS), 7(1), 12–18. Retrieved from https://www.ijars.net/articles/PDF/2436/37410_CE.pdf